A pilot study of agreement between noninvasive thermometers and the core temperature of postoperative cardiothoracic surgical patients

Published:September 18, 2022DOI:https://doi.org/10.1016/j.aucc.2022.08.008

      Abstract

      Objective

      Reliable and accurate temperature assessment is fundamental for clinical monitoring; noninvasive thermometers of various designs are widely used in intensive care units, sometimes without a specific assessment of their suitability and interchangeability. This study evaluated agreement of four noninvasive thermometers with a pulmonary artery catheter temperature.

      Methods

      This prospective method comparison study was conducted in an Australian adult intensive care unit. One hundred postoperative adult cardiothoracic surgery patients who had a pulmonary artery catheter (Edwards Lifescience) in situ were identified. The temperature reading from the pulmonary artery catheter was compared to contemporaneous measurements returned by four different thermometers—temporal Artery (TA, Technimed), Per Axilla (Axilla, Welch Allyn), Tympanic (Tymp, Covidien), and the NexTemp® (NEXT, Medical Indicators [used per axilla]). The time required to obtain each noninvasive temperature measurement was recorded.

      Results

      Agreements between each noninvasive temperature and the pulmonary artery catheter standard were assessed using summary statistics and the Bland–Altman method comparison approach. A clinically acceptable maximum difference from the standard was defined as ±0.5 °C. Temperature agreement with the pulmonary artery standard (mean difference °C [95% limits of agreement °C]) was greatest for Tymp (−0.20 [−0.92 to 0.52]), intermediate for AXILLA (−0.37 [−1.3 to 0.59]) and NEXT (−0.71 [−1.7 to 0.27]), and least for TA (−0.60 [−2.0 to 0.81]). The proportion of measurements within ±0.5 °C of the standard were TYMP (81%), AXILLA (63%), TA (45%), and NEXT (30%). The time to obtain measurements varied, with the Tymp and TA estimates immediate, the AXILLA a mean of 40 s (standard deviation = 11 s), while NEXT results were at the manufacturer-recommended 3-min point.

      Conclusions

      Tympanic thermometers showed closest agreement with the pulmonary artery standard. Deviations by more than 0.5 °C from that standard were relatively common with all noninvasive devices.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Australian Critical Care
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Fullerton J.N.
        • Thompson K.
        • Shetty A.
        • Iredell J.R.
        • Lander H.
        • Myburgh J.A.
        New sepsis definition changes incidence of sepsis in the intensive care unit.
        Crit Care Resusc. 2017; 19: 26
        • Jefferies S.
        • Weatherall M.
        • Young P.
        • Beasley R.
        A systematic review of the accuracy of peripheral thermometry in estimating core temperatures among febrile critically ill patients.
        Crit Care Resusc. 2011 Sep; 13: 194-199
        • Ramgopal S.
        • Horvat C.M.
        • Adler M.D.
        Association of triage hypothermia with in-hospital mortality among patients in the emergency department with suspected sepsis.
        J Crit Care. 2020 Dec 1; 60: 27-31
        • Niven D.J.
        • Laupland K.B.
        • Tabah A.
        • Vesin A.
        • Rello J.
        • Koulenti D.
        • et al.
        Diagnosis and management of temperature abnormality in ICUs: a EUROBACT investigators' survey.
        Crit Care. 2013; 17: R289
        • Niven D.J.
        • Gaudet J.E.
        • Laupland K.B.
        • Mrklas K.J.
        • Roberts D.J.
        • Stelfox H.T.
        Accuracy of peripheral thermometers for estimating temperature: a systematic review and meta-analysis.
        Ann Intern Med. 2015 Nov 17; 163: 768-777
        • Cutuli S.L.
        • See E.J.
        • Osawa E.A.
        • Ancona P.
        • Marshall D.
        • Eastwood G.M.
        • et al.
        Accuracy of non-invasive body temperature measurement methods in adult patients admitted to the intensive care unit: a systematic review and meta-analysis.
        Crit Care Resusc. 2021 Mar 1; 23: 6-13
        • Cutuli S.L.
        • Osawa E.A.
        • Eyeington C.T.
        • Proimos H.
        • Canet E.
        • Young H.
        • et al.
        Accuracy of non-invasive body temperature measurement methods in critically ill patients: a prospective, bicentric, observational study.
        Crit Care Resusc. 2021 Sep 6; 23: 346-353
        • Lavrakas P.
        Encyclopedia of survey research methods.
        ([Internet]. 2455 Teller Road, 91320) Sage Publications, Inc., Thousand Oaks, California, USA2008 ([cited 2022 Mar 29]. Available from:)
        • Harris P.A.
        • Taylor R.
        • Thielke R.
        • Payne J.
        • Gonzalez N.
        • Conde J.G.
        Research electronic data capture (REDCap)—a metadata-driven methodology and workflow process for providing translational research informatics support.
        J Biomed Inf. 2009 Apr 1; 42: 377-381
      1. Medical Indicators. NexTemp® “How to Use” Celsius Presentation [Internet]. Hamilton, NJ: Medical Indicators; [cited 2021 Jun 27]. Available from: https://www.medicalindicators.com/wp-content/uploads/2020/03/MII-NexTemp-How-to-Use-Slides-C.pdf.

        • Welch Allyn
        SureTemp® plus 692 electronic thermometer.
        ([Internet]) Welch Allyn, Skaneateles Falls, NY2018 ([cited 2021 Jun 27]. Available from:)
        • Visiofocus
        How to use.
        (Internet) Visofocus, Vedano Olona, Varese2013 ([cited 2021 Jun 27]. Available from)
        • Covidien
        Operating manual Genius 2 tympanic thermometer and base.
        (Internet) Covidien, Mansfield, MA2011 ([cited 2021 Jun 27]. Available from:)
        • Launey Y.
        • Larmet R.
        • Nesseler N.
        • Malledant Y.
        • Palpacuer C.
        • Seguin P.
        The accuracy of temperature measurements provided by the Edwards lifesciences pulmonary artery catheter.
        Anesth Analg. 2016 May; 122: 1480-1483
      2. Medical Indicators. NexTemp® general product literature [Internet]. Hamilton, NJ: Medical Indicators; [cited 2021 Jun 27]. Available from: https://www.medicalindicators.com/wp-content/uploads/2020/03/NexTemp-Thermometer-General-Literature.pdf.

        • Visofocus
        Specifications.
        (Internet) Visofocus, Vedano Olona, Varese2013 ([cited 2021 Jun 27]. Available from)
        • Bland J.M.
        • Altman D.G.
        Measuring agreement in method comparison studies.
        Stat Methods Med Res. 1999 Jun; 8: 135-160
        • Doménech J.
        Passing-bablok & bland-altman methods: user-written command agree for Stata.
        (Internet) Universitat Autònoma de Barcelona, Barcelona: Bellaterra2019 (Available from:)
        • StataCorp
        Stata statistical software: release. vol. 17. StataCorp LLC, College Station, TX2021
        • Lawson L.
        • Bridges E.J.
        • Ballou I.
        • Eraker R.
        • Greco S.
        • Shively J.
        • et al.
        Accuracy and precision of noninvasive temperature measurement in adult intensive care patients.
        Am J Crit Care. 2007 Sep; 16: 485-496
        • Hernandez J.M.
        • Upadhye S.
        Do peripheral thermometers accurately correlate to core body temperature?.
        Ann Emerg Med. 2016 Nov; 68: 562-563
        • Bijur P.E.
        • Shah P.D.
        • Esses D.
        Temperature measurement in the adult emergency department: oral, tympanic membrane and temporal artery temperatures versus rectal temperature.
        Emerg Med J. 2016 Dec 1; 33: 843
        • Moran J.L.
        • Peter J.V.
        • Solomon P.J.
        • Grealy B.
        • Smith T.
        • Ashforth W.
        • et al.
        Tympanic temperature measurements: are they reliable in the critically ill? A clinical study of measures of agreement.
        Crit Care Med. 2007 Jan; 35: 155-164
        • Eshraghi Y.
        • Nasr V.
        • Parra-Sanchez I.
        • Van Duren A.
        • Botham M.
        • Santoscoy T.
        • et al.
        An evaluation of a zero-heat-flux cutaneous thermometer in cardiac surgical patients.
        Anesth Analg. 2014 Sep; 119: 543-549
        • Sastre J.A.
        • Pascual M.J.
        • López T.
        Evaluation of the novel non-invasive zero-heat-flux TcoreTM thermometer in cardiac surgical patients.
        J Clin Monit Comput. 2019 Feb; 33: 165-172
        • Janke D.
        • Kagelmann N.
        • Storm C.
        • Maggioni M.A.
        • Kienast C.
        • Gunga H.C.
        • et al.
        Measuring core body temperature using a non-invasive, disposable double-sensor during targeted temperature management in post-cardiac arrest patients.
        Front Med. 2021 May 5; 8666908