Comparison of dynamic measurements of pulse contour with pulsed heat continuous cardiac output in postoperative cardiac surgical patients

Intensive Care, Prince of Wales Hospital, Barker St., Randwick, NSW 2031, Australia

Summary 

Cardiac output (CO) can be measured using bolus thermodilution via a pulmonary artery catheter (PAC) and as continuous cardiac output (CCO), using pulsed heat thermodilution. Pulse contour cardiac output (PCCO) measures continuous CO by analysis of the arterial waveform after calibration with thermodilution CO. The Pulsion Medical Systems (PiCCO system) achieves this by transpulmonary aortic thermodilution (TDtpa). There is uncertainty regarding the agreement between TDtpa, CCO, and PCCO CO measurements in situations of rapid haemodynamic changes. We studied the agreement of the measures by comparing digital recordings of cardiac index (CI) determined by PCCO and CCO (PCCI and CCI, respectively) made during periods of haemodynamic instability.

After ethics committee approval we studied four post-coronary artery bypass graft patients, in the immediate postoperative period. Each patient had a 7.5F CCO catheter (Edwards Lifesciences) and a 5F, 20cm PCCO femoral artery catheter. Digital recordings were obtained for the first 12–18 postoperative hours. Six epochs of instability were identified in the first two to three postoperative hours, and at the commencement of inotropic or vasoactive drugs. Notable features, despite frequent PCCO calibrations, were the marked difference of PCCI compared to CCI. In contradistinction, they tracked very closely during a period of stability.

Limitations of both methods were noted. Whilst PCCO responded to rapid change, it developed significant error during haemodynmamic instability and requires frequent recalibration. CCO on the other hand has a considerable time lag in responding to changes in CO. The way a monitor measures CO must be taken into account when using the data in clinical management.

Keywords: Cardiac output, Cardiac index, Cardiac surgery, Pulse contour cardiac output, Continuous cardiac output, Measurement techniques

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1036-7314(06)00005-1

doi:10.1016/j.aucc.2006.11.003

Australian Critical Care
Volume 20, Issue 1 , Pages 27-32, February 2007